The systems are designed to provide pulmonary resuscitation for infants, but the affected products may have reversed oxygen and air wall inlet fittings. This issue could reverse the oxygen concentrations and render the settings inaccurate. The product may have adverse health consequences, including death.
GE recommended that biomedical technicians or respiratory therapists check the systems and verify if the air fitting is properly installed. All affected units will need repairs.
More Articles on Anesthesia:
Study: Pain From Minor Procedures High, Not Always Well Managed
FDA Issues Device Quality Concern Notice to Hospira
ASA Opposes Proposed FDA Changes to Opioid Labels
At the Becker’s 32nd Annual Meeting: The Business and Operations of ASCs, taking place October 29-31 in Chicago, ASC leaders, surgeons and healthcare executives will explore strategies to drive growth, enhance operational performance, navigate reimbursement challenges and prepare for the future of ambulatory surgery. Apply for complimentary registration now.